HemBänkinstrumentKalibratorer/NormalerResistansGuildline 7334 Serien AC/DC Precision Air Resistance Standards

Guildline 7334 Serien AC/DC Precision Air Resistance Standards

Guildline 7334 Serien AC/DC Precision Air Resistance Standards
 
Klicka på bild för att förstora
 
 
 
 
 
 
 
 
 

  • AC/DC Error < 0.8 ppm at 1kHz
  • Stability < 2.5 ppm/Year
  • Temperature Coefficient < 0.2 ppm/°C
  • Oil or Air Baths not Required
  • Resistance Range 1 Ω to 10kΩ
  • Either Binding Posts or BPO style Connectors are available
  • Compact and Ruggedized
  • Nominal Accuracy < 2 ppm
  • Low Temperature Coefficients
  • Voltage Hysteresis < 0.1 ppm
  • Report of Calibration Traceable to NIST, NPL UK or INMS/NRCC Included
  • Guard and Shield Compliant
  • Operating Range 18 °C to 28 °C
  • Special Values Available On Request
 

 

 

Begär offert

 




 

captcha

 

GUILDLINE INSTRUMENTS 7334 SERIES of Resistance
Standards are designed as very high stability calibration
laboratory standards for high accuracy resistance calibration in air, without the need for stabilization in a temperature controlled bath.

 

These standards are suitable for both AC andDC applications.
They can be used as working standards or highly reliable and rugged transportable transfer standards.

They are extremely useful for the calibration of resistance ranges of multi-functioncalibrators and high accuracy DVMs, as well as being used in more classical standards and calibration in resistance and temperature measurements.

 

 

 
Ladda ner datablad »
 
 
 
 
Instrumentcenter Sweden AB, Box 233, 611 25 Nyköping - Sweden. | Tel: 0155-26 70 30, Fax: 0155-26 78 30
Leveransadress: Folkungavägen 4, 611 34 Nyköping